Skip to content

Unsafe health and safety: sphygmomanometer cuffs are not interchangeable (22 November 2021)

Summary

Unknown to its hypertension specialists, a major teaching hospital changed the cuffs on its sphygmomanometers from manufacturer-validated to a uniform washable alternative, in line with ‘Health and Safety’ concerns surrounding potential cross-contamination between patients. When clinic doctors suspected serious under-reading with the new cuffs, a systematic comparison was undertaken in 54 patients using two UM-101 sphygmomanometers: one using the original, manufacturer-supplied cuff and the other with the washable replacement. The study confirmed an average under-reading of 8±10/5±5 mm Hg using the washable cuff, and a third of patients with poorly controlled hypertension were considered normotensive, after using this cuff. The UM-101 sphygmomanometers have now been re-fitted with the original cuffs. Sphygmomanometer cuffs are not interchangeable between devices and a modicum of common sense should be shown to prevent changes made in the name of Health and Safety from having the opposite effect to that intended.

Unsafe health and safety: sphygmomanometer cuffs are not interchangeable (22 November 2021) https://www.nature.com/articles/jhh201251

User Feedback

Recommended Comments

There are no comments to display.

Create an account or sign in to comment

Registered address: Patient Safety Learning, China Works SB203, 100 Black Prince Road Vauxhall, London, SE1 7SJ

Account

Navigation

Search

Search

Configure browser push notifications

Chrome (Android)
  1. Tap the lock icon next to the address bar.
  2. Tap Permissions → Notifications.
  3. Adjust your preference.
Chrome (Desktop)
  1. Click the padlock icon in the address bar.
  2. Select Site settings.
  3. Find Notifications and adjust your preference.