National Patient Safety Alert: Depleted batteries in intraosseous injectors (5 December 2019)

Summary

A National Patient Safety Alert has been issued around the risk of depleted batteries in intraosseous injectors.

The intraosseous (IO) route (that is, through the bone marrow) is used to access the venous system when intravenous access is not possible to administer medicines or fluids, often in emergency situations, including cardiopulmonary resuscitation. IO access is most commonly achieved using a battery-powered injector. As the battery is sealed within the device and cannot be recharged or replaced, the first sign a battery may be depleted is in some circumstances when it does not work.

The alert asks providers to replace any battery-powered IO devices that do not have a battery power indicator light with ones with a display that shows how much power is remaining. Where IO devices with a battery power indicator are used, providers are asked to take steps to regularly check these devices to ensure sufficient battery power remains so the devices are always ready and available.

National Patient Safety Alert: Depleted batteries in intraosseous injectors (5 December 2019) https://www.england.nhs.uk/wp-content/uploads/2019/12/FINAL_Depleted_Batteries_in_IO_Devices_NatPSA_…

User Feedback

Recommended Comments

There are no comments to display.

Create an account or sign in to comment

Registered address: Patient Safety Learning, China Works SB203, 100 Black Prince Road Vauxhall, London, SE1 7SJ

Account

Navigation

Search

Search

Configure browser push notifications

Chrome (Android)
  1. Tap the lock icon next to the address bar.
  2. Tap Permissions → Notifications.
  3. Adjust your preference.
Chrome (Desktop)
  1. Click the padlock icon in the address bar.
  2. Select Site settings.
  3. Find Notifications and adjust your preference.